Customization: | Available |
---|---|
Warranty: | 1 Year |
Max Diameter of Workpiece: | Customized |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Computer Controlled Low-temperature Automatic Charpy Impact Testing Machine
1. Application:
ITM-300L type impact testing machine is used for measuring the metal material resistance performance under dynamic load, in order to determine the material properties under dynamic load. It's a necessary testing machine, not only used in the metallurgy, machine manufacture etc areas, but also used for the science research.
2. Standards:
ISO 148-1 -Metallic materials - CHARPY pendulum impact test.
JIS Z 2242 -Method for CHARPY pendulum impact test of metallic materials.
EN 10045-1 -Metallic materials-CHARPY impact test.
ASTM E23 (CHARPY) -Standard test method for notched bar impact testing of metallic materials.
GOST 9454-78 -Impact bending test method at low, room and high temperatures.
Technical Specifications:
Model | ITM-300L | ||
Display type | Computer control | ||
Max. impact energy | 300J | ||
Impact Energy | 150J | 300J | |
Value Of Every Grid | 2J | 1J | |
Moment Of Pendulum | 80.3848N·m | 160.7696N·m | |
The distance between the pendulum shaft and impact point | 750mm | ||
Impact speed | 5.24m/s | ||
Raised angle | 150° | ||
Standard span | 40+0.2mm | ||
Round angle of the jaw | R1-1.5mm | ||
Round angle of impact edge | R2-2.5mm,R8±0.05(special order)mm | ||
Angle accuracy | 0.1° | ||
Standard specimen dimension | 10mm×10(7.5/5)mm×55mm | ||
Thickness of impact knife | 16mm | ||
Dip angle of supporting surface of supports | 11°±1° | ||
Impact edge Angle | 30°±1° | ||
Impact edge curvature radius |
2mm impact edge | 2-2.5 | |
8mm impact blade | 8±0.05 | ||
8mm impact shoulder Angle radius | 0.1-1 | ||
8mm impact blade width | 4 ±0.05 | ||
Impact blade width | 10-18mm | ||
Impact knife thickness | 16mm | ||
Meet the sample size | 10×10x55 mm | ||
Cooling way | Compressor | ||
Capacity of sample box | 20 | ||
Range of low temperature | room tempeature~-60°c | ||
Range of low temperature | fluctuation ±0.5°C grades 1°C | ||
Speed of sending specimen | ≤2S | ||
Power supply | 3phs, 380V, 50Hz Or 220V,60Hz | ||
Measured parameters | |||
Angle measurement | photoelectric encoder | ||
Angle accuracy | 0.1° | ||
Dial energy resolution | ≥0.5J | ||
computer screen display energy resolution | 0.01J | ||
Energy display mode | Dial , microcomputer |
Main Unit | Quantity |
Host Part | |
Main frame e |
1 PC |
Dial | 1 PC |
Pendulum(150J/300J) | 2 PCS |
Motor 370W | 1 PC |
Electromagnetic clutch |
1 PC |
Swing Driving gear | 1 PC |
Pendant swing mechanism | 1 PC |
Semi-automatic protective cover |
1 PC |
Full-closed enclosure(Optional) | 1 PC |
Low temperature refrigeration and sample delivery positioning system | |
Compressor | 2 PC |
Filter dryer | 1 PC |
oil-water separator |
1 PC |
Condensing apparatus |
2 PC |
High precision intelligent temperature controller | 1 PC |
Stainless steel heater | 1 PC |
Automatic hanging and swinging device | 1 PC |
Sample cylinder | 1 PC |
Locating cylinder | 1 PC |
Computer and HP Printer | 1 PC |
Electrical and measurement and control systems. |
|
Programmable controller for PLC |
1 PC |
PC Testing-software | 1 PC |
Computer and HP Printer | 1 PC |
Manual control box |
1 PC |
Test fixtures and Tools |
|
Pendulum(150J/300J) | 2 PCS |
Pendulum striking edge 2mm | 2 PCS |
Internal Hexagonal Wrench (S=12mm) | 1 PC |
Span Specimen Plate (L=40mm) | 1 PC |
Specimen Centering Plate (L=40mm)(U, V) | 1 PC |
Stripping Attachment | 1 PC |